2
Atomic weight: 4.003
Isotope abundance: 4 (100%)
Oxidation states: 0
AA Flame Type: Not detectable by AA
3
Atomic weight: 6.941
Chemical form in solution: Li+(aq) (large effective radius due to hydration sphere)
Isotope abundance: 6 (7.59%); 7 (92.41%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Direct dissolution | Dissolves very rapidly in water |
| Ores | Sodium carbonate fusion + acid dissolution | - Fusion with Na₂CO₃ in platinum crucible - Dissolve melt in HCl - Li blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion | - H₂SO₄ + H₂O₂ digestion or - HNO₃ + H₂SO₄ + HClO₄ decomposition |
| Part No. | Type | Description |
|---|---|---|
| 5610103000 | Coded | Lithium (Li) coded single element hollow cathode lamp |
| 5610124900 | Uncoded | Lithium (Li) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
4
Atomic weight: 9.012
Chemical form in solution: Be2+(H2O)42+
Isotope abundance: 9 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 5-10% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in diluted H₂SO₄ |
| BeO | Acid digestion or fusion | - Boiling HNO₃, HCl, or H₂SO₄ or - KHSO₄ fusion |
| Ores | Acid digestion or fusion | - H₂SO₄ + HF digestion or - Carbonate fusion in platinum crucible - Li blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion or dry ashing | - H₂SO₄ + H₂O₂ or HNO₃ + H₂SO₄ + HClO₄ or - Dry ash + BeO procedure dissolution |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8565 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610100500 | Coded | Beryllium (Be) coded single element hollow cathode lamp |
| 5610122400 | Uncoded | Beryllium (Be) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
9
Atomic weight: 18.998
Isotope abundance: 19 (100%)
Oxidation states: -1
AA Flame Type: Not detectable by AA
10
Atomic weight: 20.179
Isotope abundance: 20 (90.48%); 21 (0.27%); 22 (9.25%)
Oxidation states: 0
AA Flame Type: Not detectable by AA
11
Atomic weight: 22.98977
Chemical form in solution: Na+(H2O)n+
Isotope abundance: 23 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Direct dissolution | Dissolves very rapidly in water |
| Ores | Lithium carbonate fusion + acid dissolution | - Fusion with Na₂CO₃ in graphite crucible - Dissolve melt in HCl - Na blank in Li₂CO₃ must be low |
| Organic Matrices | Wet digestion | - H₂SO₄ + H₂O₂ digestion or - HNO₃ + H₂SO₄ + HClO₄ decomposition |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-411 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610105300 | Coded | Sodium (Na) coded single element hollow cathode lamp |
| 5610127300 | Uncoded | Sodium (Na) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610107000 | Coded | Na - K, Coded HC Lamp 1/pk |
| 5610107090 | Coded | Na - K Coded HC Lamp, Certified, 1/pk |
12
Atomic Weight: 24.305
Chemical form in solution: Mg(H2O)6+2
Isotope abundance: 24 (78.99%)*; 25 (10%); 26 (11.01%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-10% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in diluted HNO₃ |
| Oxide | Acid dissolution | Readily soluble in compatible aqueous acidic solutions |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl - Li blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion or dry ashing | - H₂SO₄ + H₂O₂ or HNO₃ + H₂SO₄ + HClO₄ or - Dry ash + dissolve in dilute HCl (do not heat to avoid SiO₂ precipitation) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-412 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610103200 | Coded | Magnesium (Mg) coded single element hollow cathode lamp |
| 5610125100 | Uncoded | Magnesium (Mg) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108800 | Coded | Al - Ca - Mg, Coded HC Lamp, 1/pk |
| 5610133600 | Coded | Al-Ca-Mg, Coded UltrAA HC Lamp, 1/pk |
13
Atomic Weight: 26.98154
Chemical form in solution: Al(H2O)6+3
Isotope abundance: 27 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in HCl / HNO₃ |
| Oxide (α-Al₂O₃) | Fusion | Requires Na₂CO₃ fusion in Pt crucible |
| Oxide (γ-Al₂O₃) | Acid dissolution | Soluble in acids such as HCl |
| Ores | Fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or ashing + acid dissolution | - H₂SO₄ / H₂O₂ digestion or HNO₃ / H₂SO₄ / HClO₄ decomposition - Dry ash, then dissolve in dilute HCl |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8561 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610100100 | Coded | Aluminum (Al) coded single element hollow cathode lamp |
| 5610122000 | Uncoded | Aluminum (Al) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108800 | Coded | Al - Ca - Mg, Coded HC Lamp, 1/pk |
| 5610133600 | Coded | Al-Ca-Mg, Coded UltrAA HC Lamp, 1/pk |
17
Atomic weight: 35.453
Isotope abundance: 35 (75.76%)*; 37 (24.24%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +6, +5, +4, +3, +1, -1, -2
AA Flame Type: Not detectable by AA
18
Atomic weight: 39.948
Isotope abundance: 36 (0.34%); 38 (0.06%); 40 (99.60%)
Oxidation states: 0
AA Flame Type: Not detectable by AA
19
Atomic weight: 39.0983
Chemical form in solution: K+(aq)
Isotope abundance: 39 (93.26%)*; 40 (0.01%); 41 (6.73%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Direct dissolution | Dissolves very rapidly in water |
| Ores | Sodium carbonate fusion + acid dissolution | - Fusion with Na₂CO₃ in platinum crucible - Dissolve melt in HCl - K blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion | - H₂SO₄ + H₂O₂ digestion |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-419 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610104200 | Coded | Potassium (K) coded single element hollow cathode lamp |
| 5610126200 | Uncoded | Potassium (K) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610107000 | Coded | Na - K, Coded HC Lamp 1/pk |
| 5610107090 | Coded | Na - K Coded HC Lamp, Certified, 1/pk |
20
Atomic weight: 40.078
Chemical form in solution: Ca(H2O)6+2
Isotope abundance: 40 (96.941%); 42 (0.647%); 43 (0.135%)*; 44 (2.086%)*; 46 (0.004%); 48 (0.187%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-10% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in diluted HNO₃ |
| Ores | Fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl |
| Organic Matrices | Dry ashing + acid dissolution | - Dry ash and dissolve in dilute HCl (do not heat to avoid SiO₂ precipitation) |
| Compounds | Direct acid dissolution or fusion | - Oxide, hydroxide, carbonate, phosphate, fluoride: soluble in %
HCl or HNO₃ - Sulfates, silicates, complex compounds: require Na₂CO₃ fusion + HCl/water dissolution |
| Trace Analysis | - | Contamination control is critical when analyzing Ca at trace levels |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-420 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610101000 | Coded | Calcium (Ca) coded single element hollow cathode lamp |
| 5610122900 | Uncoded | Calcium (Ca) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108800 | Coded | Al - Ca - Mg, Coded HC Lamp, 1/pk |
| 5610133600 | Coded | Al-Ca-Mg, Coded UltrAA HC Lamp, 1/pk |
21
Atomic weight: 44.95591
Chemical form in solution: Sc(H2O)6+2
Isotope abundance: 45 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 5-10% HNO3/LDPE container. Small atomic radius increases hydrolysis requiring higher acid levels than other Rare Earths.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ - Aqua regia or HNO₃ + HClO₄ + H₂SO₄ digestions also suitable (caution with HClO₄) |
| Part No. | Type | Description |
|---|---|---|
| 5610126900 | Uncoded | Scandium (Sc) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
22
Atomic weight: 47.88
Chemical form in solution: Ti(F)6-2
Isotope abundance: 46 (8.25%); 47 (7.44%)*; 48 (73.72%); 49 (5.41%); 50 (5.18%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals) as the Ti(F)6-2 for months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the Ti(F)6-2 chemically stable for years in 2-5% HNO3/trace HF in an LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in H₂O / HF — CAUTION: Ti powder reacts violently |
| Oxide (anatase/rutile) | Heated acid dissolution | Dissolve by heating in 1:1:1 H₂O / HF / H₂SO₄ |
| Oxide (anatase/rutile) | Fusion | Fuse in platinum crucible with K₂S₂O₇ |
| Ores | Fusion + acid dissolution | - Fuse with KF + K₂S₂O₇ in platinum crucible - Omit KF if no silica present |
| Organic Matrices | Ashing + acid dissolution or fusion | - Dry ash at 450°C in platinum - Dissolve by heating with 1:1:1 H₂O / HF / H₂SO₄ - If present as plastic pigment (brookite form), fuse ash with pyrosulfate |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-422 | 10 µg/mL | 0.5% HNO3 with 0.1% HF | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610106200 | Coded | Titanium (Ti) coded single element hollow cathode lamp |
| 5610128200 | Uncoded | Titanium (Ti) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
23
Atomic weight: 50.9415
Chemical form in solution: H2V10O284-
Isotope abundance: 50 (0.25%); 51 (99.75%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +5, +4, +3, +2, +1, 0, -1
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Fusion | Fuse with NaOH or KOH in Ni crucible or with Na₂CO₃ / KNO₃ |
| Oxide | Acid dissolution | - V₂O₃: use HCl - V₂O₄: use HCl or HNO₃ - V₂O₅: use concentrated acids |
| Ores | Fusion + water extraction | - Fuse with Na₂CO₃ / KNO₃ in Pt crucible - Caution: nitrates attack Pt - Extract fuseate with H₂O |
| Organic Matrices | Ashing + acid dissolution | - Ash at 450°C, then dissolve as for V₂O₅ (i.e., in concentrated acids) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8585 | 10 µg/mL | 5% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610106500 | Coded | Vanadium (V) coded single element hollow cathode lamp |
| 5610128500 | Uncoded | Vanadium (V) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
24
Atomic weight: 51.9961
Chemical form in solution: Cr(H2O)63+
Isotope abundance: 50 (4.345%); 52 (83.79%)*; 53 (9.50%); 54 (2.365%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2, +1, 0, -1, -2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HCl |
| Oxides / Ores | Multiple fusion methods (A–D) | Very difficult to dissolve; common procedures: A. KHSO₄ fusion → hot KCl extract → Na₂CO₃ + KClO₃ (3:1) B. NaKSO₄ + NaF fusion (2:1) C. Magnesia or lime + Na or K carbonates fusion (4:1) D. Na₂O₂ or NaOH + KNO₃ or NaOH + Na₂O₂ fusion • Use Ni, Fe, Cu, or Ag crucibles for D • Pt crucibles suitable for A, B, C |
| Organic Matrices | Ashing + fusion or acid digestion | - Ash at 450°C then apply one of the fusion methods (A–D) - For non-oxide samples, H₂SO₄ + H₂O₂ digestion may be suitable |
| Part No. | Type | Description |
|---|---|---|
| 5610101200 | Coded | Chromium (Cr) coded single element hollow cathode lamp |
| 5610123100 | Uncoded | Chromium (Cr) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610109500 | Coded | Ag-Cr-Cu-Fe-Ni, Coded HC Lamp, 1/pk |
| 5610134900 | Coded | Ag-Cr-Cu-Fe-Ni Coded UltrAA HC Lamp 1/pk |
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
25
Atomic weight: 54.9380
Chemical form in solution: Mn(H2O)62+
Isotope abundance: 55 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +6, +5, +4, +3, +2, +1, 0, -1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in dilute acids |
| Oxides | Acid dissolution | Soluble in dilute acids |
| Ores | Acid digestion with silica removal | - Dissolve with HCl - If silica is present, add HF, then fume off silica by adding H₂SO₄ and heating to SO₃ fumes (dense white fumes) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8574 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610103300 | Coded | Manganese (Mn) coded single element hollow cathode lamp |
| 5610125200 | Uncoded | Manganese (Mn) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610133700 | High intensity UltrAA | Manganese (Mn) high intensity UltrAA coded single element hollow cathode lamp |
| Part No. | Type | Description |
|---|---|---|
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
| 5610109600 | Coded | Cu - Fe - Mn - Zn, Coded HC Lamp, 1/pk |
| 5610135000 | Coded | Cu-Fe-Mn-Zn, Coded UltrAA HC Lamp, 1/pk |
26
Atomic weight: 55.847
Chemical form in solution: Fe(H2O)63+
Isotope abundance: 54 (5.85%); 56 (91.75%)*; 57 (2.12%); 58 (0.28%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2, +1, 0
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HCl |
| Oxides | Fusion or acid dissolution | - If high-temperature oxide, use Na₂CO₃ fusion in Pt crucible
followed by HCl dissolution - Otherwise, dissolve directly in dilute HCl |
| Ores | Fusion + acid dissolution | Follow oxide fusion method only: Na₂CO₃ fusion in Pt crucible → dissolve in HCl |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-426 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610102700 | Coded | Iron (Fe) coded single element hollow cathode lamp |
| 5610124600 | Uncoded | Iron (Fe) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610108600 | High intensity UltrAA | Iron (Fe) high intensity UltrAA coded single element hollow cathode lamp |
| Part No. | Type | Description |
|---|---|---|
| 5610109500 | Coded | Ag-Cr-Cu-Fe-Ni, Coded HC Lamp, 1/pk |
| 5610134900 | Coded | Ag-Cr-Cu-Fe-Ni Coded UltrAA HC Lamp 1/pk |
| 5610135300 | Uncoded | As-Cu-Fe, Uncoded UltrAA HC Lamp, 1/pk |
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610109600 | Coded | Cu - Fe - Mn - Zn, Coded HC Lamp, 1/pk |
| 5610135000 | Coded | Cu-Fe-Mn-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610109700 | Coded | Cu - Fe - Si - Zn, Coded HC Lamp, 1/pk |
| 5610135100 | Coded | Cu-Fe-Si-Zn, Coded UltrAA HC Lamp, 1/pk |
27
Atomic weight: 58.9332
Chemical form in solution: Co(H2O)62+
Isotope abundance: 59 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +5, +4, +3, +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 8500-6947 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610101300 | Coded | Cobalt (Co) coded single element hollow cathode lamp |
| 5610123200 | Uncoded | Cobalt (Co) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610134100 | High intensity UltrAA | Cobalt (Co) high intensity UltrAA coded single element hollow cathode lamp |
| Part No. | Type | Description |
|---|---|---|
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
| 5610135200 | Coded | Co-Mo-Pb-Zn, Coded UltrAA HC Lamp, 1/pk |
28
Atomic weight: 58.6934
Chemical form in solution: Ni(H2O)62+
Isotope abundance: 58 (68.08%); 60 (26.22%)*; 61 (1.14%); 62 (3.63%); 64 (0.93%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +4, +3, +2, +1, 0, -1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8576 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610103700 | Coded | Nickel (Ni) coded single element hollow cathode lamp |
| 5610125600 | Uncoded | Nickel (Ni) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610108500 | High intensity UltrAA | Nickel (Ni) high intensity UltrAA coded single element hollow cathode lamp |
| Part No. | Type | Description |
|---|---|---|
| 5610109500 | Coded | Ag-Cr-Cu-Fe-Ni, Coded HC Lamp, 1/pk |
| 5610134900 | Coded | Ag-Cr-Cu-Fe-Ni Coded UltrAA HC Lamp 1/pk |
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
| 5610135400 | Uncoded | Ni - Se, Uncoded UltrAA HC Lamp, 1/pk |
29
Atomic weight: 63.546
Chemical form in solution: Cu(H2O)62+
Isotope abundance: 63 (69.15%)*; 65 (30.85%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8569 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610101400 | Coded | Copper (Cu) coded single element hollow cathode lamp |
| 5610123300 | Uncoded | Copper (Cu) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610109500 | Coded | Ag-Cr-Cu-Fe-Ni, Coded HC Lamp, 1/pk |
| 5610134900 | Coded | Ag-Cr-Cu-Fe-Ni Coded UltrAA HC Lamp 1/pk |
| 5610135300 | Uncoded | As-Cu-Fe, Uncoded UltrAA HC Lamp, 1/pk |
| 5610134500 | Coded | Co-Cr-Cu-Fe-Mn-Ni Coded UltrAA Lamp 1/pk |
| 5610107600 | Coded | Co-Cr-Cu-Fe-Mn-Ni, Coded HC Lamp, 1/pk |
| 5610129200 | Uncoded | Co-Cr-Cu-Fe-Mn-Ni, Uncoded HC Lamp, 1/pk |
| 5610109600 | Coded | Cu - Fe - Mn - Zn, Coded HC Lamp, 1/pk |
| 5610135000 | Coded | Cu-Fe-Mn-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610109700 | Coded | Cu - Fe - Si - Zn, Coded HC Lamp, 1/pk |
| 5610135100 | Coded | Cu-Fe-Si-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610119200 | Coded | Copper - Zinc Cu-Zn, Coded HC Lamp, 1/pk |
| 5610134600 | Coded | Cu - Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610129300 | Uncoded | Cu - Zn, Uncoded HC Lamp, 1/pk |
30
Atomic weight: 65.38
Chemical form in solution: Zn(OH)(aq)+
Isotope abundance: 64 (49.17%); 66 (27.73%)*; 67 (4.04%); 68 (18.45%); 70 (0.61%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Organic Matrices | Ashing or wet digestion | - Dry ash at 450°C, then dissolve in HCl - Alternatively, use H₂SO₄ / H₂O₂ digestion |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8587 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610106800 | Coded | Zinc (Zn) coded single element hollow cathode lamp |
| 5610128800 | Uncoded | Zinc (Zn) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108700 | Coded | Ag - Cd - Pb - Zn, Coded HC Lamp, 1/pk |
| 5610108900 | Coded | Ag-Cd-Pb-Zn Coded UltrAA HC Lamp, 1/pk |
| 5610135200 | Coded | Co-Mo-Pb-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610109600 | Coded | Cu - Fe - Mn - Zn, Coded HC Lamp, 1/pk |
| 5610135000 | Coded | Cu-Fe-Mn-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610109700 | Coded | Cu - Fe - Si - Zn, Coded HC Lamp, 1/pk |
| 5610135100 | Coded | Cu-Fe-Si-Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610119200 | Coded | Copper - Zinc Cu-Zn, Coded HC Lamp, 1/pk |
| 5610134600 | Coded | Cu - Zn, Coded UltrAA HC Lamp, 1/pk |
| 5610129300 | Uncoded | Cu - Zn, Uncoded HC Lamp, 1/pk |
31
Atomic weight: 69.723
Chemical form in solution: Ga(H2O)6+3
Isotope abundance: 69 (60.11%); 71 (39.89%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in HCl / HNO₃ |
| Oxide (Ga₂O₃) | Fusion | Requires Na₂CO₃ fusion in Pt crucible |
| Ores | Fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or ashing + acid dissolution | - H₂SO₄ / H₂O₂ digestion or HNO₃ / H₂SO₄ / HClO₄ decomposition - Dry ash, then dissolve in dilute HCl |
| Part No. | Type | Description |
|---|---|---|
| 5610123800 | Uncoded | Gallium (Ga) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
35
Atomic weight: 79.904
Isotope abundance: 79 (50.69%)*; 81 (49.31%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +5, +4, +3, +1, -1
AA Flame Type: Not detectable by AA
36
Atomic weight: 83.8
Isotope abundance: 78 (0.35%); 80 (2.29%); 82 (11.59%); 83 (11.50%); 84 (56.99%); 86 (17.28%)
Oxidation states: +2
AA Flame Type: Not detectable by AA
37
Atomic weight: 85.4678
Chemical form in solution: Rb+(aq)
Isotope abundance: 85 (72.17%)*; 87 (27.83%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 0.1-1% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Direct dissolution | Dissolves very rapidly in water |
| Ores | Sodium carbonate fusion + acid dissolution | - Fusion with Na₂CO₃ in platinum crucible - Dissolve melt in HCl - Li blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion | - H₂SO₄ + H₂O₂ digestion or - HNO₃ + H₂SO₄ + HClO₄ decomposition |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8440 | 10,000 µg/mL | 5% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610126600 | Uncoded | Rubidium (Rb) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
38
Atomic weight: 87.62
Chemical form in solution: Sr(H2O)6+2
Isotope abundance: 84 (0.56%); 86 (9.86%); 87 (7.00%); 88 (82.58%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-3.5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in diluted HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl - Li blank in Na₂CO₃ must be low |
| Organic Matrices | Dry ashing + acid dissolution | - Dry ash and dissolve in dilute HCl (do not heat to avoid SiO₂ precipitation) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8581 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610105400 | Coded | Strontium (Sr) coded single element hollow cathode lamp |
| 5610127400 | Uncoded | Strontium (Sr) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
39
Atomic weight: 88.9059
Chemical form in solution: Y(OH)(H2O)x+2
Isotope abundance: 89 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610128700 | Uncoded | Yttrium (Y) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
40
Atomic weight: 91.224
Chemical form in solution: Zr(F)6-2
Isotope abundance: 90 (51.45%)*; 91 (11.22%); 92 (17.15%); 94 (17.38%); 96 (2.80%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3, +2, 0, -2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the Zr(F)6-2 + Zr(OH)4F2-2 for months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the Zr(F)6-2 chemically stable for years in 2-5% HNO3/trace HF in an LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in H₂O / HF / HNO₃ |
| Oxide | Fusion | Unlike TiO₂, ZrO₂ is best fused using: - Na₂O₂ in Ni crucible - Na₂CO₃ in Pt crucible - Borax in Pt crucible |
| Organic Matrices | Ashing + fusion + acid dissolution | - Dry ash at 450°C in Pt crucible - Fuse with Na₂CO₃ - Dissolve in HF / HNO₃ / H₂O |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8327 | 1,000 µg/mL | 5% HNO3 | 100 mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610106900 | Coded | Zirconium (Zr) coded single element hollow cathode lamp |
| 5610128900 | Uncoded | Zirconium (Zr) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
41
Atomic weight: 92.9064
Chemical form in solution: NbOF5-2
Isotope abundance: 93 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +5, +4, +3, +2, +1, -1, -3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the NbOF5-2 for 5 months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the NbOF5-2 chemically stable for years in 2-5% HNO3/trace HF in an LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HF / HNO₃ |
| Oxide | Fusion | Very resistant to all acids (including HF); requires fusion
with: - K₂S₂O₇ - KOH - Na₂CO₃ |
| Organic Matrices | Ashing + fusion | - Dry ash at 450°C in Pt crucible - Fuse with Na₂CO₃ or K₂S₂O₇ |
| Part No. | Type | Description |
|---|---|---|
| 5610103800 | Coded | Niobium (Nb) coded single element hollow cathode lamp |
| 5610125700 | Uncoded | Niobium (Nb) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
42
Atomic weight: 95.96
Chemical form in solution: MoO4-2 (chem. form as received)
Isotope abundance: 92 (14.53%); 94 (9.15%); 95 (15.84%)*; 96 (16.67%); 97 (9.60%); 98 (24.39%); 100 (9.82%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the MoOF5-2 for months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the MoO4-2 chemically stable for years in 1% NH4OH in a LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HF / HNO₃ or hot dilute HCl |
| Oxide | Acid or base dissolution | Soluble in HF or NH₄OH |
| Organic Matrices | Ashing + acid dissolution | - Dry ash at 450°C in Pt crucible - Dissolve oxide with HF or HCl |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8297 | 1,000 µg/mL | 1% HN4OH | 100 mL | CRM |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-442 | 10 µg/mL | 0.5% NH4OH | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610103500 | Coded | Molybdenum (Mo) coded single element hollow cathode lamp |
| 5610125400 | Uncoded | Molybdenum (Mo) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610135200 | Coded | Co-Mo-Pb-Zn, Coded UltrAA HC Lamp, 1/pk |
43
Atomic weight: (98)
Oxidation states: +7, +6, +5, +4, 0, -1
AA Flame Type: Not detectable by AA
44
Atomic weight: 101.07
Chemical form in solution: RuCl6
Isotope abundance: 96 (5.54%); 98 (1.87%); 99 (12.76%); 100 (12.60%); 101 (17.06%)*; 102 (31.55%); 104 (18.62%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +8, +7, +6, +5, +4, +3, +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 10% HCl/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Fusion | Fuse with KOH / KNO₃ in Ag crucible |
| Oxides | Fusion | Fuse with KOH / KNO₃ in Ag crucible |
| Ores | Fusion | Same as oxides: KOH / KNO₃ fusion in Ag crucible |
| Alloys | Fusion | Same as oxides: KOH / KNO₃ fusion in Ag crucible |
| Organic Matrices | Fusion + acid dissolution | RuO₄ is volatile – avoid aggressive acidic
oxidation Preferred method: KOH / KNO₃ fusion, dissolve fuseate in HCl |
| Part No. | Type | Description |
|---|---|---|
| 5610126700 | Uncoded | Ruthenium (Ru) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
45
Atomic weight: 102.91
Chemical form in solution: RhCl6-3
Isotope abundance: 103 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2, +1, 0, -1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 10% HCl/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | High-temperature acid digestion | Requires elevated temperature with aqua regia or HCl + Cl₂(gas) |
| Ores | Sequential acid digestion | - Initial digestion with HF / H₂SO₄ |
| Platinum Scrap | Acid digestion | Digestion with aqua regia |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 8500-6945 | 10 µg/mL | 2% HCl | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610104500 | Coded | Rhodium (Rh) coded single element hollow cathode lamp |
| 5610126500 | Uncoded | Rhodium (Rh) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
46
Atomic weight: 106.42
Chemical form in solution: Pd(H2O)62+
Isotope abundance: 102 (1.02%); 104 (11.14%); 105 (22.33%)*; 106 (27.33%); 108 (26.46%); 110 (11.72%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +2, 0
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels. 2 ppb Pd is stable for 1 day in 1% HNO3/LDPE container. 10 ppb is stable for 3 days in 1% HNO3/LDPE container. 100 ppb is stable for ≥ 5 months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ or aqua regia |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610104000 | Coded | Palladium (Pd) coded single element hollow cathode lamp |
| 5610125900 | Uncoded | Palladium (Pd) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610135800 | High intensity UltrAA | Palladium (Pd) high intensity UltrAA coded single element hollow cathode lamp |
47
Atomic weight: 107.8682
Chemical form in solution: Ag(H2O)6+
Isotope abundance: 107 (51.84%)*; 109 (48.16%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for 75+ days when mixed with equivalent levels of all other elements including the precious metals (where chloride is present) when in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HNO₃ |
| Ores | Acid digestion | Digest with concentrated HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8580 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610105200 | Coded | Silver (Ag) coded single element hollow cathode lamp |
| 5610127200 | Uncoded | Silver (Ag) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108700 | Coded | Ag - Cd - Pb - Zn, Coded HC Lamp, 1/pk |
| 5610108900 | Coded | Ag-Cd-Pb-Zn Coded UltrAA HC Lamp, 1/pk |
| 5610109500 | Coded | Ag-Cr-Cu-Fe-Ni, Coded HC Lamp, 1/pk |
| 5610134900 | Coded | Ag-Cr-Cu-Fe-Ni Coded UltrAA HC Lamp 1/pk |
48
Atomic weight: 112.41
Chemical form in solution: Cd2(OH)(aq)3+ and Cd(OH)(aq)+
Isotope abundance: 106 (1.25%); 108 (0.89%); 110 (12.49%); 111 (12.80%)*; 112 (24.13%); 113 (12.22%); 114 (28.73%); 116 (7.49%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides | Acid dissolution | Soluble in HCl or HNO₃ |
| Ores | Acid digestion + fuming + filtration | - Dissolve in HCl / HNO₃, then bring to fumes with H₂SO₄ - After adding water, filter off silica and lead sulfate |
| Organic Matrices | Ashing or wet digestion | - Dry ash at 450°C, dissolve in HCl - Alternatively, use H₂SO₄ / H₂O₂ digestion |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8567 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610100800 | Coded | Cadmium (Cd) coded single element hollow cathode lamp |
| 5610122700 | Uncoded | Cadmium (Cd) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| Part No. | Type | Description |
|---|---|---|
| 5610108700 | Coded | Ag - Cd - Pb - Zn, Coded HC Lamp, 1/pk |
| 5610108900 | Coded | Ag-Cd-Pb-Zn Coded UltrAA HC Lamp, 1/pk |
49
Atomic weight: 114.82
Chemical form in solution: In(H2O)6+3
Isotope abundance: 113 (4.29%); 115 (95.71%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in HCl / HNO₃ |
| Oxide | Acid dissolution | Soluble in mineral acids |
| Ores | Fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or ashing + acid dissolution | - H₂SO₄ / H₂O₂ digestion or Dry ash and dissolve in dilute HCl |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8284 | 1,000 µg/mL | 5% HNO3 | 100 mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610124400 | Uncoded | Indium (In) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
50
Atomic weight: 118.710
Chemical form in solution: Sn(OH)xFy2-
Isotope abundance: 112 (0.97%); 114 (0.66%); 115 (0.34%); 116 (14.54%); 117 (7.68%); 118 (24.22%)*; 119 (8.59%); 120 (32.58%); 122 (4.63%); 124 (5.79%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the Sn(OH)xFy2- for 1 year in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the Sn(OH)xFy2- are chemically stable for years in 2-5% HNO3/trace HF in a LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HF / HNO₃ or HCl |
| Oxide – SnO | Acid dissolution | Soluble in HCl |
| Oxide – SnO₂ | Fusion + aqueous dissolution | Very resistant to all acids including HF Fuse with equal parts Na₂CO₃ + S Thiostannate formed is soluble in H₂O or dilute acids |
| Alloys | Sequential acid digestion | - Treat 0.1 g alloy with 10 mL conc. H₂SO₄, boil until
disintegration - Add 100 mL O₂-free water + 50 mL conc. HCl or Transfer to plastic, add 1 mL HF, warm gently to dissolve |
| Organic Matrices | Specialized handling recommended | Volatility and SnO₂ precipitation are common issues. Please
contact technical support at: 📧 chem-standards-support@agilent.com 📞 (800) 227-9770 for tailored guidance based on your specific sample type https://community.agilent.com/technical/consumables/ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8583 | 10 µg/mL | 5% HCl | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610106100 | Coded | Tin (Sn) coded single element hollow cathode lamp |
| 5610128100 | Uncoded | Tin (Sn) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610133900 | High intensity UltrAA | Tin (Sn) high intensity UltrAA coded single element hollow cathode lamp |
53
Atomic weight: 126.905
Isotope abundance: 127 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +5, +3, +1, 0, -1
AA Flame Type: Not detectable by AA
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Elemental Iodine (I₂) | Controlled oxidation or alkaline dissolution | - Volatile and easily lost during heating or open digestions - Dissolve in alkaline iodide (e.g., NaOH + NaI) to stabilize as I⁻ - Avoid acidic conditions unless using a closed system |
| Oxides & Iodates | Acid or reductive dissolution | - Soluble in dilute HCl or reducing agents (e.g., thiosulfate for IO₃⁻) |
| Ores / Minerals | Alkaline fusion or reductive leaching | - Use Na₂CO₃ / NaOH fusion in Pt or Ni crucible for insoluble
iodides - Alternatively, leach with NaHSO₃ or Na₂S₂O₃ under reducing conditions |
| Organic Matrices | Mild alkaline digestion or combustion | - Dry ash avoided due to I₂ volatility |
| Special Notes | Volatility | - I₂ is volatile and can be lost during ashing, heating, or acidic
digestion - Closed-vessel digestion or alkaline stabilization strongly recommended |
54
Atomic weight: 131.29
Isotope abundance: 124 (0.09%); 126 (0.09%); 128 (1.92%); 129 (26.40%); 130 (4.07%); 131 (21.23%); 132 (26.91%); 134 (10.44%); 136 (8.86%)
Oxidation states: +8, +6, +4, +2
AA Flame Type: Not detectable by AA
55
Atomic weight: 132.9054
Chemical form in solution: Cs+(aq)
Isotope abundance: 133 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 0.1% HNO3/LDPE container. Will crystallize out of higher (~ 5%) levels of HNO3 at > 1000 µg/mL.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Direct dissolution | Dissolves very rapidly in water |
| Ores | Sodium carbonate fusion + acid dissolution | - Fusion with Na₂CO₃ in platinum crucible - Dissolve melt in HCl - Cs blank in Na₂CO₃ must be low |
| Organic Matrices | Wet digestion | - H₂SO₄ + H₂O₂ digestion or - HNO₃ + H₂SO₄ + HClO₄ decomposition |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8274 | 1,000 µg/mL | 5% HNO3 | 100 mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610122800 | Uncoded | Cesium (Cs) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
56
Atomic weight: 137.33
Chemical form in solution: Ba(H2O)6+2
Isotope abundance: 130 (0.11%); 132 (0.10%); 134 (2.42%); 135 (6.59%); 136 (7.85%); 137 (11.23%)*; 138 (71.70%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-3.5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in diluted HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in platinum crucible - Dissolve melt in HCl - If sulfate is present, use HCl + tartaric acid to prevent BaSO₄ precipitation |
| Organic Matrices | Dry ashing + acid dissolution | - Dry ash and dissolve in dilute HCl (do not heat to avoid SiO₂ precipitation) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8564 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610100400 | Coded | Barium (Ba) coded single element hollow cathode lamp |
| 5610122300 | Uncoded | Barium (Ba) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
72
Atomic weight: 178.49
Chemical form in solution: Hf(F)6-2
Isotope abundance: 174 (0.16%); 176 (5.26%); 177 (18.60%); 178 (27.28%)*; 179 (13.62%); 180 (35.08%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable alone or mixed with all other metals that are at comparable levels as the Hf(F)6-2 + Hf(OH)4F2-2 for months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the Hf(F)6-2 chemically stable for years in 2-5% HNO3/trace HF in an LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in H₂O / HF / HNO₃ |
| Oxide | Fusion | Unlike TiO₂, HfO₂ is best fused using:
- Na₂O₂ in Ni crucible - Na₂CO₃ in Pt crucible - Borax in Pt crucible |
| Organic Matrices | Ashing + fusion + acid dissolution | - Dry ash at 450°C in Pt crucible - Fuse with Na₂CO₃ - Dissolve in HF / HNO₃ / H₂O |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-172 | 10,000 µg/mL | 0.2% HF | 125mL | RM |
| Part No. | Type | Description |
|---|---|---|
| 5610102200 | Coded | Hafnium (Hf) coded single element hollow cathode lamp |
| 5610124100 | Uncoded | Hafnium (Hf) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
73
Atomic weight: 180.9479
Chemical form in solution: TaOF6-3
Isotope abundance: 180 (0.01%); 181 (99.99%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +5, +4, +3, +1, -1, -3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the TaOF6-3, for 2 months at the 2-10 ppb level in 1% HNO3/LDPE container and for 5 months at the 100 ppb level under same conditions. 1-10,000 ppm single element solutions as the TaOF6-3 are chemically stable for years in 2-5% HNO3/trace HF in an LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HF / HNO₃ |
| Oxide | Fusion | Very resistant to all acids (including HF); requires fusion
with: - K₂S₂O₇ - KOH - Na₂CO₃ |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610105500 | Coded | Tantalum (Ta) coded single element hollow cathode lamp |
| 5610127500 | Uncoded | Tantalum (Ta) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
74
Atomic weight: 183.85
Chemical form in solution: WOF5-2 (chem. form as received)
Isotope abundance: 180 (0.12%); 182 (26.50%)*; 183 (14.31%); 184 (30.64%); 186 (28.43%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2, 0, -1, -2, -4
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable (alone or mixed with all other metals that are at comparable levels) as the [WOF5]-2 for months in 1% HNO3/LDPE container. 1-10,000 ppm single element solutions as the [WOF5]-2 chemically stable for years in 1% HF in a LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HF / HNO₃ |
| Oxide | Acid or base dissolution | Soluble in HF or NH₄OH |
| Organic Matrices | Ashing + acid dissolution | - Dry ash at 450°C in Pt crucible - Dissolve oxide with HF |
| Part No. | Type | Description |
|---|---|---|
| 5610128300 | Uncoded | Tungsten (W) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
75
Atomic weight: 186.207
Chemical form in solution: ReO4-
Isotope abundance: 185 (37.40%)*; 187 (62.60%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +6, +5, +4, +3, +2, +1, 0, -1, -3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 1-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxides / Ores | Fusion | Fuse in Pt crucible with Na₂CO₃ |
| Organic Matrices | Ashing + controlled acid digestion | - All acid digestions risk volatile perrhenic acid loss - Use reflux condenser if applying HNO₃/HClO₄ or H₂SO₄/H₂O₂ - Preferred method: mix sample with Na₂CO₃, ash in Pt crucible starting at 450°C, increase up to 900°C to fuse alumino-silicates, etc. |
| Part No. | Type | Description |
|---|---|---|
| 5610104400 | Coded | Rhenium (Re) coded single element hollow cathode lamp |
| 5610126400 | Uncoded | Rhenium (Re) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
76
Atomic weight: 190.2
Chemical form in solution: OsCl62-
Isotope abundance: 184 (0.02%); 186 (1.59%); 187 (1.96%); 188 (13.24%); 189 (16.15%)*; 190 (26.26%); 192 (40.78%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +8, +7, +6, +5, +4, +3, +2, +1, 0, -2
AA Flame Type: N20/Acetylene
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Oxides | Fusion + aqueous dissolution | - Fuse with KOH / KNO₃ in Ag crucible - Dissolve only in water — avoid acids |
| Ores | Same as oxides | Use KOH / KNO₃ fusion in Ag crucible, dissolve in water |
| Organic Matrices | Fusion + aqueous dissolution | - OsO₄ is volatile — use KOH / KNO₃ fusion, dissolve fuseate in
water - Avoid acidic oxidizing digestion - APDC has been used to help stabilize Os but needs further validation |
| Special Notes | Volatility & interference | - OsO₄ enhances nebulization efficiency, causing falsely high
results - Only make dilutions in HCl - Avoid HNO₃ - Caustic nitrate fusion products should be diluted only in water |
77
Atomic weight: 192.22
Chemical form in solution: IrCl6-2
Isotope abundance: 191 (37.3%); 193 (62.7%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3, +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels. 2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 10% HCl/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | High-temperature acid digestion | Requires elevated temperature with aqua regia or HCl + Cl₂(gas) |
| Ores | Sequential acid digestion | - Digest with HF / H₂SO₄ - Follow with aqua regia digestion |
| Platinum Scrap | Acid digestion | Digestion with aqua regia |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8570 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610102600 | Coded | Iridium (Ir) coded single element hollow cathode lamp |
| 5610124500 | Uncoded | Iridium (Ir) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
78
Atomic weight: 195.08
Chemical form in solution: Pt(Cl)62-
Isotope abundance: 190 (0.012%); 192 (0.782%); 194 (32.86%); 195 (33.78%)*; 196 (25.21%); 198 (7.356%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +2, 0
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-10 ppb Pt is stable for 2 months in 1% HNO3/LDPE container. 100 ppb is stable for 5 months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 10% HCl/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in aqua regia |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8577 | 10 µg/mL | 5% HCl | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610104100 | Coded | Platinum (Pt) coded single element hollow cathode lamp |
| 5610126100 | Uncoded | Platinum (Pt) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610135900 | High intensity UltrAA | Platinum (Pt) high intensity UltrAA coded single element hollow cathode lamp |
79
Atomic weight: 196.9665
Chemical form in solution: Au(Cl)63-
Isotope abundance: 197 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +7, +5, +3, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels. 2-10 ppb Au is stable for ≤ 1 day maximum in 1% HNO3/LDPE container. 100 ppb is stable for ≤ 2 days maximum in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 10% HCl/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in aqua regia |
| Oxides | Acid dissolution | Soluble in HCl |
| Ores | Acid digestion | Dissolve in a mixture of HCl / HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 8500-7000 | 100 µg/mL | 2% HCl | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610102100 | Coded | Gold (Au) coded single element hollow cathode lamp |
| 5610124000 | Uncoded | Gold (Au) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610109000 | High intensity UltrAA | Gold (Au) high intensity UltrAA coded single element hollow cathode lamp |
80
Atomic weight: 200.59
Chemical form in solution: Hg(OH)(aq)+
Isotope abundance: 196 (0.15%); 198 (9.97%); 199 (16.87%); 200 (23.10%); 201 (13.18%)*; 202 (29.86%)*; 204 (6.87%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +2, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels—stable in 10% HNO3 packaged in borosilicate glass; NOT stable in 1% HNO3/ LDPE container. 1-100 ppm levels stable in 7% HNO3 packaged in borosilicate glass. 1000-10,000 ppm solutions are chemically stable for years in 5-10% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| HgO (Oxide) | Acid dissolution | Soluble in HNO₃ |
| Ores & Organic Matrices | Specialized preparation required | Preparations for Hg are prone to error. Please contact technical
support at: 📧 chem-standards-support@agilent.com 📞 (800) 227-9770 for tailored guidance based on your specific sample type https://community.agilent.com/technical/consumables/ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8575 | 10 µg/mL | 5% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610103400 | Coded | Mercury (Hg) coded single element hollow cathode lamp |
| 5610125300 | Uncoded | Mercury (Hg) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
81
Atomic weight: 204.383
Chemical form in solution: Tl(H2O)6+
Isotope abundance: 203 (29.52%); 205 (70.48%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +1
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in HNO₃, forming primarily the Tl⁺ (monovalent) ion |
| Oxide | Dissolution (depends on oxidation state) | - Thallous oxide: readily soluble in water - Thallic oxide: requires high acid concentration |
| Ores | Fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or ashing + acid dissolution | - H₂SO₄ / H₂O₂ digestion or Dry ash, then dissolve in HCl |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8316 | 1,000 µg/mL | 5% HNO3 | 100 mL | CRM |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-481 | 10 µg/mL | 0.5% HNO3 | 125mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610105800 | Coded | Thallium (Tl) coded single element hollow cathode lamp |
| 5610127800 | Uncoded | Thallium (Tl) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610108400 | High intensity UltrAA | Thallium (Tl) high intensity UltrAA coded single element hollow cathode lamp |
82
Atomic weight: 207.2
Chemical form in solution: Pb(H2O)6+2
Isotope abundance: 204 (1.4%); 206 (24.1%); 207 (22.1%); 208 (52.4%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +2
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Best dissolved in 1:1 H₂O / HNO₃ |
| Oxides | Acid dissolution | Most Pb oxides are soluble in HNO₃ PbO₂ requires HCl or HF |
| Ores & Alloys | Acid digestion | Best attacked with 1:1 H₂O / HNO₃ |
| Organic Matrices | Ashing + acid dissolution | Dry ash, dissolve in dilute HCl Do not heat to avoid SiO₂ precipitation |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8571 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610102900 | Coded | Lead (Pb) coded single element hollow cathode lamp |
| 5610124800 | Uncoded | Lead (Pb) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610108200 | High intensity UltrAA | Lead (Pb) high intensity UltrAA coded single element hollow cathode lamp |
| Part No. | Type | Description |
|---|---|---|
| 5610108700 | Coded | Ag - Cd - Pb - Zn, Coded HC Lamp, 1/pk |
| 5610108900 | Coded | Ag-Cd-Pb-Zn Coded UltrAA HC Lamp, 1/pk |
| 5610135200 | Coded | Co-Mo-Pb-Zn, Coded UltrAA HC Lamp, 1/pk |
83
Atomic weight: 208.9804
Chemical form in solution: Bi(O)(H2O)x+
Isotope abundance: 209 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +5, +3, +1, -3
AA Flame Type: Air/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 5-7% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in HNO₃ |
| Oxide | Acid dissolution | Soluble in HNO₃ |
| Alloys | Mixed acid digestion | Dissolve in conc. 4:1 HCl / HNO₃; heating may be required |
| Organic Matrices | Ashing or wet digestion | - Dry ash at 450°C, dissolve in HNO₃ - Or use hot conc. H₂SO₄ + H₂O₂ (added carefully dropwise) until clear |
| Part No. | Type | Description |
|---|---|---|
| 5610100600 | Coded | Bismuth (Bi) coded single element hollow cathode lamp |
| 5610122500 | Uncoded | Bismuth (Bi) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
| 5610134200 | High intensity UltrAA | Bismuth (Bi) high intensity UltrAA coded single element hollow cathode lamp |
85
Atomic weight: (210)
Oxidation states: +3, +1, -1
AA Flame Type: Not detectable by AA
86
Atomic weight: (222)
Oxidation states: +2
AA Flame Type: Not detectable by AA
87
Atomic weight: (223)
Oxidation states: +1
AA Flame Type: Not detectable by AA
88
Atomic weight: 226.025
Oxidation states: +2
AA Flame Type: Not detectable by AA
104
Atomic weight: (267)
AA Flame Type: Not detectable by AA
105
Atomic weight: (268)
AA Flame Type: Not detectable by AA
106
Atomic weight: (269)
AA Flame Type: Not detectable by AA
107
Atomic weight: (270)
AA Flame Type: Not detectable by AA
108
Atomic weight: (277)
AA Flame Type: Not detectable by AA
109
Atomic weight: (278)
AA Flame Type: Not detectable by AA
110
Atomic weight: (281)
AA Flame Type: Not detectable by AA
111
Atomic weight: (282)
AA Flame Type: Not detectable by AA
112
Atomic weight: (285)
AA Flame Type: Not detectable by AA
113
Atomic weight: (284)
AA Flame Type: Not detectable by AA
114
Atomic weight: (289)
AA Flame Type: Not detectable by AA
115
Atomic weight: (288)
AA Flame Type: Not detectable by AA
116
Atomic weight: (293)
AA Flame Type: Not detectable by AA
117
Atomic weight: (294)
AA Flame Type: Not detectable by AA
118
Atomic weight: (294)
AA Flame Type: Not detectable by AA
57
Atomic weight: 138.9055
Chemical form in solution: La(OH)y(H2O)x+3-y
Isotope abundance: 138 (0.09%); 139 (99.91%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610124700 | Uncoded | Lanthanum (La) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
58
Atomic weight: 140.12
Chemical form in solution: Ce(OH)y(H2O)x+4-y
Isotope abundance: 136 (0.19%); 138 (0.25%); 140 (88.45%)*; 142 (11.11%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution |
59
Atomic weight: 140.9077
Chemical form in solution: Pr(OH)y(H2O)x+3-y
Isotope abundance: 141 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
60
Atomic weight: 144.24
Chemical form in solution: Nd(OH)y(H2O)x+3-y
Isotope abundance: 142 (27.2%); 143 (12.2%); 144 (23.8%); 145 (8.3%); 146 (17.2%)*; 148 (5.7%); 150 (5.6%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610103600 | Coded | Neodymium (Nd) coded single element hollow cathode lamp |
| 5610125500 | Uncoded | Neodymium (Nd) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
61
Atomic weight: (145)
Oxidation states: +3
AA Flame Type: Not detectable by AA
62
Atomic weight: 150.36
Chemical form in solution: Sm(OH)y(H2O)x+3-y
Isotope abundance: 144 (3.07%); 147 (14.99%)*; 148 (11.24%); 149 (13.82%); 150 (7.38%); 152 (26.75%); 154 (22.75%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8515 | 1,000 µg/mL | 5% HNO3 | 100mL | CRM |
| Part No. | Type | Description |
|---|---|---|
| 5610126800 | Uncoded | Samarium (Sa) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
63
Atomic weight: 151.96
Chemical form in solution: Eu(OH)y(H2O)x+3-y
Isotope abundance: 151 (47.81%); 153 (52.19%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610123600 | Uncoded | Europium (Eu) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
64
Atomic weight: 157.25
Chemical form in solution: Gd(OH)x(H2O)y+3-x
Isotope abundance: 152 (0.2%); 154 (2.18%); 155 (14.80%); 156 (20.47%); 157 (15.65%)*; 158 (24.84%); 160 (21.86%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part No. | Type | Description |
|---|---|---|
| 5610123700 | Uncoded | Gadolinium (Gd) uncoded single element hollow cathode lamp, for all AAS instruments with octal base connector |
65
Atomic weight: 158.9254
Chemical form in solution: Tb(OH)x(H2O)y+3-x
Isotope abundance: 159 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible - Dissolve melt in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8216 | 10,000 µg/mL | 500mL | CRM |
66
Atomic weight: 162.50
Chemical form in solution: Dy(OH)x(H2O)y+3-x
Isotope abundance: 156 (0.06%); 158 (0.09%); 160 (2.33%); 161 (18.89%); 162 (25.47%); 163 (24.90%)*; 164 (28.26%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
67
Atomic weight: 164.9304
Chemical form in solution: Ho(OH)x(H2O)y+3-x
Isotope abundance: 165 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
68
Atomic weight: 167.26
Chemical form in solution: Er(OH)x(H2O)y+3-x
Isotope abundance: 162 (0.14%); 164 (1.60%); 166 (33.50%)*; 167 (22.87%); 168 (26.98%); 170 (14.91%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
69
Atomic weight: 168.9342
Chemical form in solution: Tm(OH)x(H2O)y+3-x
Isotope abundance: 169 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3, +2
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash, then dissolve in 1:1 H₂O / HCl or HNO₃ |
70
Atomic weight: 173.054
Chemical form in solution: Yb(OH)x(H2O)y+3-x
Isotope abundance: 168 (0.12%); 170 (2.98%); 171 (14.09%); 172 (21.68%)*; 173 (16.10%); 174 (32.03%); 176 (13.00%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
71
Atomic weight: 174.9668
Chemical form in solution: Lu(OH)x(H2O)y+3-x
Isotope abundance: 175 (97.40%)*; 176 (2.60%)
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +3
AA Flame Type: N20/Acetylene
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in acids |
| Oxide | Heated acid dissolution | Dissolved by heating in H₂O / HNO₃ |
| Ores | Carbonate fusion + acid dissolution | - Carbonate fusion in Pt crucible, followed by dissolution in HCl |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8573 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
89
Atomic weight: 227.028
Oxidation states: +3
AA Flame Type: Not detectable by AA
90
Atomic weight: 232.0381
Chemical form in solution: Th(OH)3+ and Th(OH)22+
Isotope abundance: 232 (100%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +4, +3, +2
AA Flame Type: Not detectable by AA
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Soluble in aqua regia |
| Oxide | Acid dissolution (limited) | Heated oxide is not soluble in most acids; only soluble in hot conc. H₂SO₄ |
| Ores | Fusion + sequential acid treatment | - Fuse with Na₂O₂ at 480 ± 20°C for 7 min - Cool and treat with cold H₂O, let stand to disintegrate - Acidify with HCl + 25 mL excess HCl - Filter any residue (Whatman No. 42), dry and ignite to 1000°C in Pt crucible - Ash treated with H₂SO₄ / HF and fumed - If residue remains, apply peroxide fusion as above |
| Organic Matrices | Wet digestion or acid dissolution | - Dry ash and dissolve in 1:1 H₂O / HCl or HNO₃ - Aqua regia or HNO₃ + HClO₄ + H₂SO₄ digestions also suitable (caution with HClO₄) |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-190 | 10,000 µg/mL | 4% HNO3 | 125mL | RM |
91
Atomic weight: 231.036
Oxidation states: +5, +4, +3
AA Flame Type: Not detectable by AA
92
Atomic weight: 238.0289
Chemical form in solution: UO22+ (uranyl)
Isotope abundance: 234 (0.005%); 235 (0.720%); 238 (99.274%)*
Highlighted isotope is the recommended mass number for ICP-MS analysis
Oxidation states: +6, +5, +4, +3
AA Flame Type: Not detectable by AA
Store at room temperature (15° to 30°C)
2-100 ppb levels stable for months in 1% HNO3/LDPE container. 1-10,000 ppm solutions chemically stable for years in 2-5% HNO3/LDPE container.
| Sample Type | Preparation Type | Notes |
|---|---|---|
| Metal | Acid dissolution | Dissolves rapidly in HCl and HNO₃ |
| Oxide | Acid dissolution | Soluble in HNO₃ |
| Ores | Acid digestion + silica removal | - Digest 1 g ore in 30 mL of 1:1 HNO₃ for 1–2 hours - Bring to fumes with conc. H₂SO₄ - Remove silica insolubles by filtration |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| ICP-192 | 10,000 µg/mL | 4% HNO3 | 125mL | RM |
| Part Number | Concentration | Matrix | Volume | Quality Level (ISO 17034) |
|---|---|---|---|---|
| 5190-8584 | 10 µg/mL | 2% HNO3 | 100mL | CRM |
93
Atomic weight: 237.048
Oxidation states: +7, +6, +5, +4, +3, +2
AA Flame Type: Not detectable by AA
94
Atomic weight: (244)
Oxidation states: (+7), +6, +5, +4, +3, +2
AA Flame Type: Not detectable by AA
95
Atomic weight: (243)
Oxidation states: +6, +5, +4, +3, +2
AA Flame Type: Not detectable by AA
96
Atomic weight: (247)
Oxidation states: +4, +3, +2
AA Flame Type: Not detectable by AA
97
Atomic weight: (247)
Oxidation states: +4, +3, +2
AA Flame Type: Not detectable by AA
98
Atomic weight: (251)
Oxidation states: +4, +3, +2
AA Flame Type: Not detectable by AA
99
Atomic weight: (254)
Oxidation states: (+4), +3, (+2)
AA Flame Type: Not detectable by AA
100
Atomic weight: (257)
Oxidation states: +3, (+2)
AA Flame Type: Not detectable by AA
101
Atomic weight: (258)
Oxidation states: +3, +2
AA Flame Type: Not detectable by AA
102
Atomic weight: (259)
Oxidation states: +3, +2
AA Flame Type: Not detectable by AA
103
Atomic weight: (262)
Oxidation states: +3
AA Flame Type: Not detectable by AA
Accurately detecting heavy metals and other elements in your samples relies on the quality of your chemical standards and supplies. All Agilent reference and certified reference materials are rigorously tested and manufactured using ISO 17025 and 17034 accreditations.
Download a printable copy with a full relative isotopic abundance table.
For specific agency method mixes and kits, please click here. And click here to view our full chemical standards portfolio.
CRM – Certified Reference Material = A standards whose property values are sufficiently homogeneous, stable and well established to be used for maintaining or monitoring measurement processes. Typically used for inter-laboratory testing.
RM – Reference Materials = Standards that have been tested to be sufficiently homogeneous and stable with respect to one or more specified properties, which has been established to be fit for its intended use in a measurement process. Standards are manufactured in ISO 17034 accredited facility.
Accurate, productive elemental analysis demands sample preparation that is straightforward, cost effective, and compatible with your analytical goals and methods. From filtration and purification to solid phase extraction, Agilent makes delivering accurate results easier. Learn more
Vials are a critical part of your analytical workflow. Choose Agilent vials and closures and gain complete confidence in your sample containment—and your results. Learn more
ULTRA Scientific has been acquired by Agilent. That means we can now offer you a full portfolio of reliable and effective reference materials, along with easy online ordering, rapid shipping, and custom solutions. Learn more
Agilent’s range of high-quality ICP-OES supplies includes nebulizers, torches and application kits. The fully demountable torch for the Agilent 5000 Series ICP-OES instruments features a removable injector to simplify torch maintenance and allows quick changeover for diverse sample matrices. Learn more
Never compromise on AA performance. Agilent hollow cathode lamps deliver excellent stability and sensitivity, improved detection limit performance, and a longer service life. Learn more
Agilent interface cones are designed and tested on Agilent ICP-MS systems to ensure sensitivity and stability with real-world applications. They offer superior sensitivity across the mass range with the lowest background.Learn more
Productive, precise, and reliable. Agilent leads the industry with innovative AA instruments to match every need and budget—including the world’s fastest flame AA system. Learn more
Perform fast, sensitive elemental analysis. Agilent ICP-MS systems give you excellent accuracy, ease of use, and productivity. Learn more
Whether for metals analysis, elemental analysis or investigation of organic solvents, our high-quality ICP-OES systems, accessories, supplies, and ICP Expert software can help you reach your analytical goals. Learn more
See our entire portfolio of instruments. Learn more
Our lab software portfolio features a range of high-quality solutions to ensure your laboratory can make the most of every analytical system. Learn more
Avoid the time and costs of developing new methods from scratch, restoring methods after a repair, and transitioning methods to new instruments. Agilent CrossLab application support can help you confidently deploy the latest productivity and usability enhancements and ensure the best outcomes—from sample preparation to final report. Learn more
Agilent CrossLab service engineers have the skills, qualifications, and experience to manage all your instrument servicing, regardless of the manufacturer. Learn more